Use the Operand menu to select how Ansur shall calculate the limit given according to the power values defined
in the test details. The operand can be set to:
Y – absolute value,
X + Y – limit calculated as defined power value + specified limit, or
X + (X * Y%) – limit calculated as a percentage deviance from the defined power value.
So for the example below, each power performance measurement must be +/-20 W of its defined power value.